Includes IPL power supply+laser power supply+RF power supply+laser capacitor+control system+dual handle conversion+TEC cooling control;
Powerful supporting software: human-machine interaction operation screen (customizable)+backend parameter setting+one screen display of IPL, laser, RF operation interface+real-time parallel operation of IPL, laser, RF+power calibration storage and function+multi language support+online program upgrade;
Input power factor ≥ 0.99;
The whole machine is designed in accordance with the YY0505-2012 and IEC60601-1-2 standards for medical electrical equipment.
Integrating IPL/laser/RF three major energy systems, supporting dual handle conversion and TEC cooling, achieving multifunctional synchronous treatment.
Customizable human-machine interface, displaying multiple parameters on one screen, supporting multiple languages and online upgrades, with precise and convenient operation.
Power factor ≥ 0.99, in compliance with YY0505/IEC60601 standards, ensuring stable energy output and safe treatment.
| Specifications | Parameter |
|---|---|
| Input Voltage | AC220V±15%,47-63Hz AC110V±15%,47-63Hz |
| Leakage Current | <0.4mA/250AC |
| Maximum output electrical energy of laser single pulse (under the condition of 100uF capacitor) | 45J |
| IPL single pulse maximum output energy (under capacitor 20000uF condition) | 825J |
| RF Output Parameters | 1MHz |
| Temperature | Work:-15℃~45℃ Storage:-40℃~85℃ |
| Humidity | No frost or condensation |
| Safety regulations and electromagnetic compatibility | Designed in accordance with YY0505-2012 and IEC60601-1-2 standards for medical electrical equipment |
| Cooling Method | forced air cooling |
| Dimensions(mm) | 282*148*197 |
| Weight(KG) | 9.5 |
Warranty: 360 days
▶ The warranty period for power products is 12 months, calculated from the date of purchase.
▶ During the warranty period, faults that are not caused by human factors can be repaired free of charge; Malfunctions caused by human damage or excessive use require paid repairs.
▶ Before using this power supply, please read the manual carefully. We are not responsible for any damage or problems caused by improper use or failure to follow the manual
Submit this form and we will contact you soon