ND:YAG Passive Laser Power Supply System

Integrated design: including pre burning laser power supply, control system, and capacitor;

Powerful supporting software: human-machine interaction operation screen (customizable)+adjustable backend voltage parameters;

Unique high current plug and unplug device, easy to install and disassemble, saving labor costs;

Adopting new technology and modular design to ensure efficiency, stability, and anti-interference;

Fault problem location, including over temperature, water flow, pre combustion, and charging alarm prompts;

Patent design can ensure high consistency of continuous pulse output energy;

Medical beauty power supply comes standard with a protective casing, which is radiation resistant, anti bounce, and has a stable installation structure;

The whole machine is designed in accordance with the YY0505-2012 and IEC60601-1-2 standards for medical electrical equipment.

intelligent integration

Pre combustion power supply+control+capacitor three in one design, matched with customizable software, flexible parameter adjustment, intuitive and convenient operation.

High energy stability

Patent technology ensures pulse energy consistency, modular anti-interference design, and high current plug to improve maintenance efficiency.

Medical protection

Compliant with YY0505/IEC60601 standards, equipped with intelligent fault alarm and radiation resistant casing, ensuring safe and reliable treatment.

Technical Specifications

Power Parameters
ModelSpecificationsQPM500QPM2000
Input ParametersInput VoltageAC220V±15%,47 - 63Hz
AC110V±15%,47 - 63Hz
AC220V±15%,47 - 63Hz/110V
Leakage Current< 0.4mA/250AC< 0.4mA/250AC
Output ParametersOutput Power500W2000W

Maximum output energy per pulse

(Under the condition of a capacitance of 100 μ F)

50J200J
Output Frequency1 - 10HzSingle time,1HZ-5HZ
Electric pulse width<150μs<450μs
Touch Screen4.3h、8h、10.4h
TemperatureWork:-15℃ - 45℃
Storage:-40℃ - 85℃
HumidityNo frost or condensation
Safety regulations and electromagnetic compatibilityDesigned in accordance with YY0505-2012 and IEC60601-1-2 standards for medical electrical equipment
Cooling Methodforced air cooling
Dimensions (mm)270*124*124315*200*135
Weight (KG)3.14.9

Quality Assurance

Warranty: 360 days


▶ The warranty period for power products is 12 months, calculated from the date of purchase.


▶ During the warranty period, faults that are not caused by human factors can be repaired free of charge; Malfunctions caused by human damage or excessive use require paid repairs.


▶ Before using this power supply, please read the manual carefully. We are not responsible for any damage or problems caused by improper use or failure to follow the manual