Integrated design: including pre burning laser power supply, control system, and capacitor;
Powerful supporting software: human-machine interaction operation screen (customizable)+adjustable backend voltage parameters;
Unique high current plug and unplug device, easy to install and disassemble, saving labor costs;
Adopting new technology and modular design to ensure efficiency, stability, and anti-interference;
Fault problem location, including over temperature, water flow, pre combustion, and charging alarm prompts;
Patent design can ensure high consistency of continuous pulse output energy;
Medical beauty power supply comes standard with a protective casing, which is radiation resistant, anti bounce, and has a stable installation structure;
The whole machine is designed in accordance with the YY0505-2012 and IEC60601-1-2 standards for medical electrical equipment.
Pre combustion power supply+control+capacitor three in one design, matched with customizable software, flexible parameter adjustment, intuitive and convenient operation.
Patent technology ensures pulse energy consistency, modular anti-interference design, and high current plug to improve maintenance efficiency.
Compliant with YY0505/IEC60601 standards, equipped with intelligent fault alarm and radiation resistant casing, ensuring safe and reliable treatment.
| Power Parameters | |||
| ModelSpecifications | QPM500 | QPM2000 | |
| Input Parameters | Input Voltage | AC220V±15%,47 - 63Hz AC110V±15%,47 - 63Hz | AC220V±15%,47 - 63Hz/110V |
| Leakage Current | < 0.4mA/250AC | < 0.4mA/250AC | |
| Output Parameters | Output Power | 500W | 2000W |
Maximum output energy per pulse (Under the condition of a capacitance of 100 μ F) | 50J | 200J | |
| Output Frequency | 1 - 10Hz | Single time,1HZ-5HZ | |
| Electric pulse width | <150μs | <450μs | |
| Touch Screen | 4.3h、8h、10.4h | ||
| Temperature | Work:-15℃ - 45℃ Storage:-40℃ - 85℃ | ||
| Humidity | No frost or condensation | ||
| Safety regulations and electromagnetic compatibility | Designed in accordance with YY0505-2012 and IEC60601-1-2 standards for medical electrical equipment | ||
| Cooling Method | forced air cooling | ||
| Dimensions (mm) | 270*124*124 | 315*200*135 | |
| Weight (KG) | 3.1 | 4.9 | |
Warranty: 360 days
▶ The warranty period for power products is 12 months, calculated from the date of purchase.
▶ During the warranty period, faults that are not caused by human factors can be repaired free of charge; Malfunctions caused by human damage or excessive use require paid repairs.
▶ Before using this power supply, please read the manual carefully. We are not responsible for any damage or problems caused by improper use or failure to follow the manual
Submit this form and we will contact you soon